Data Loading...

Edelman Chapter 021 - Orange Coast College Flipbook PDF

3 AIUM 100mm Test Object • Dead Zoneis evaluated by imaging the pins located superficially in the phantom and determinin


233 Views
134 Downloads
FLIP PDF 192.52KB

DOWNLOAD FLIP

REPORT DMCA

Quality Assurance Quality Assurance Chapter 21

Goals • Assure proper operation of the system and it components • detect gradual changes • reduce critical down time • decrease the number of equipment produced poor quality exams • minimize repeat imaging due to system problems

Types of devices used • Doppler Phantom

• Beam Profile/slice thickness Phantom

• The routine and periodic evaluation of ultrasound systems to assure optimal image quality • QA programs include • assessment of system components • system repairs • PM • and of course the hated paper work problem of record keeping

Types of devices used • AIUM 100 mm test object

• Tissue Equivalent Phantoms

AIUM 100mm Test Object • Fluid fill with strategically placed pins • the fluid should be that of normal soft tissue • there is not attenuation properties of the fluid • designed to evaluate the accuracy and performance characteristics of the system

1

AIUM 100mm Test Object

A Dead Zone E vertical registration B Lateral resolution When

• Axial resolution evaluated when the pins are parallel to the sounds main beam axis

transducer is on oblique corner

B Axial resolution. If transducer is on top C Lateral resolution When transducer is on side C Axial resolution If transducer is on top D Horizontal Registration

• The smallest distance in which 2 pins parallel to the primary beam are displayed as separate objects

AIUM 100mm Test Object • Lateral resolution is evaluated by measuring the size of the pins located perpendicular to the sound beam • Minimum distance two side by side rods are displayed as separate entities (scan side B)

Registration accuracy

located parallel to the beam • If irregularities occur between scan and pin location causes may be system malfunction or sound speed in phantom is different than 1,450 m/s

AIUM 100mm Test Object • Electronic caliper accuracy of the system compares the system measurements against a phantom standard

AIUM 100mm Test Object • The ability of the system to echo’s in the correct position with differing orientations • Vertical depth calibration means the system places the pins at the correct depth when

AIUM 100mm Test Object

• Horizontal calibration is similar to vertical but the pins are located perpendicular to the beam

AIUM 100mm Test Object • Dead Zone is evaluated by imaging the pins located superficially in the phantom and determining the minimum depth a pin can be resolved creating a meaningful reflection

2

AIUM 100mm Test Object • Dead Zone is evaluated by imaging the pins located superficially in the phantom and determining the minimum depth a pin can be resolved creating a meaningful reflection • With a tissue equivalent phantom this zone is the shallowest depth in which uniform tissue texture appears

AIUM 100mm Test Object • Focal Zone is a point where the beam is the narrowest and the beams intensity is the greatest. • Lateral resolution is very good and the beam is narrow • array transducers are user controlled and should be carefully evaluated • dynamic receive focusing should produce narrow reflections through a wide range of depths

Uniformity • The systems ability to display similar reflectors throughout the phantom with equal brightness and regardless of the depth • The same TGC and depth should be used from test to test

Tissue Equivalent Phantoms • Mimics the features of normal soft tissue • evaluates characteristics in modern systems – – – –

gray scale tissue texture multi-focus adjustable focus found in phased array transducers

Solid and cystic uniformity • With a tissue equivalent phantom solid masses should appear uniform • cystic masses should be anechoic

Tissue Equivalent Phantoms • Phantom has the following soft tissue associations – – – –

speed of sound attenuation scattering characteristics echogenicity

• evaluates gray scale by viewing hollow and solid cysts within the phantom

3

Doppler Phantom • Used to assess characteristics of all Doppler systems including Pulsed , CW and Power • Uses motion from a string moving belt or fluid flow phantom

Sensitivity • Is the ability of the system to resolve low level echo’s and display them. • There are three types – minimum – normal – maximum

Normal Sensitivity • A setting at which all pins in the phantom can be displayed • Output power, TGC,and amplification are set to establish normals • normal sensitivity is at a higher gain than minimum. • All subsequent QA tests are then made at this normal setting • There should be no variance from test to test over time

Slice Thickness Phantoms • Used to determine the systems elevation resolution • measures the beam geometry perpendicular to the imaging plane • the phantom contains a diffuse scattering plane at an angle to the primary beam

• The imaging plane is thicker than the beam width or the pulse length so slice thickness is most likely to degrade image quality

Minimum Sensitivity • Assesses the weakest sonographic echo intensity that can be accurately displayed from the far field • Start by setting the TGC flat then increase the system gain from the minimal setting to just when the deep rods appear • There should be no variance from test to test over time

Maximum Sensitivity • A evaluation of power output and amplification with the system set to the maximum practical levels. • A tissue equivalent phantom is then imaged using these settings. • The depth of the tissue mimicking texture is determined . • The max. visualization depth is then assessed • There should be no variance from test to test over time

4